Rationale
and Discussion for Proposed Algorithm
Itemized Discussion Points
(a) Call for help (the anesthesia technician, other
clinicians)
At this point, for the hypoxic "O2" pipeline algorithm to
be initiated, there is a low FiO2 alarm and/or a low FiO2 reading that
does not match the set FiO2. The clinician should not wait for a low
SpO2 alarm to respond because if the patient is pre-oxygenated before
a gas mix-up occurs, it may be several minutes before SpO2 drops. The
clinician should be taught to set the low FiO2 alarm close to the desired
FiO2. Thus, if set FiO2 is 0.5, the low FiO2 alarm should be set at
0.4 or 0.45, not at 0.21. The clinician should trust the oxygen sensor
and SpO2 monitor and not waste precious time verifying whether the O2
sensor reads correctly in room air (this should have been already done
anyway during step 9a of the 1993 FDA pre-use check). Help should be
called because additional hands and brains are helpful in managing a
problem.
(b) Disconnect the patient from the circuit at the
elbow connector with the gas sampling line still attached to the elbow
connector
The primary responsibility of the clinician is to the patient, not
debugging the suspected fault with the machine. Therefore, isolate the
patient from the machine as soon as a machine malfunction is suspected.
If the Y-piece needs to be closed off to debug the machine, use a 3-L
reservoir bag as the "patient". The gas sampling line should
be left on the circuit when disconnecting from the patient so that in
step (f) of the proposed algorithm, FiO2 can be monitored while flushing
with O2.
(c) Ventilate (or ask for assistance to manually
ventilate) with a self-inflating resuscitation bag (SIRB or "Ambu"
bag) using room air. If the patient needs O2-enriched air, use O2 from
a stand-alone O2 cylinder.
Manual ventilation with a self inflating resuscitation bag (SIRB or
"Ambu" bag) with room air containing 21% O2 is better than
continued mechanical or manual ventilation (using the 3 L reservoir
bag) with a hypoxic gas mixture when the objective is to re-establish
oxygenation and a normal SpO2. If the patient requires O2-enriched air
during ventilation with an SIRB or Mapleson system, do not use supplemental
O2 from an auxiliary O2 flow meter or common gas outlet installed on
the anesthesia machine because these auxiliary O2 outlets will also
supply hypoxic gas if the O2 pipeline is still connected to the anesthesia
machine.
(d) Disconnect O2 pipeline - audible alarm should
sound confirming O2 pipeline disconnection
The O2 pipeline has to be disconnected for O2 to flow from the reserve
O2 cylinder. This step is in anticipation of step (e). O2 pipeline disconnection
precedes opening the O2 cylinder so that hypoxic gas inflow into the
machine is interrupted as soon as possible. Also, disconnecting the
O2 pipeline before opening the cylinder will, in most anesthesia machines,
generate an audible alarm that provides a confirmatory cue to the clinician
that the correct pipeline (O2) has been disconnected. In the heat of
the action, it may be possible that a clinician disconnects the N2O
pipeline instead of the O2 pipeline such that opening the O2 cylinder
in the ensuing step does not help. If the O2 cylinder is opened first,
there will be no audible alarm when the O2 pipeline is subsequently
disconnected. The clinician may check that the O2 pipeline pressure
gauge reads zero to verify that the O2 pipeline is disconnected.
(e) Open the O2 cylinder - audible sound should
indicate that O2 cylinder is open and not empty
The O2 cylinder is opened to provide access to the reserve O2 stored
in the O2 cylinder. An audible signal will be generated in most anesthesia
machines to provide a confirmatory cue to the clinician that the O2
cylinder has been opened and the anesthesia machine is now being supplied
with O2 from the O2 cylinder. If the O2 cylinder had been opened before
the O2 pipeline was disconnected, there would be no audible confirmatory
cue that O2 is flowing from the O2 cylinder. The clinician may also
check that the O2 cylinder pressure gauge reads more than zero to verify
that the O2 cylinder is not empty.
(f) Press the oxygen flush button until the O2 analyzer
displays rapidly increasing FiO2 values
The hypoxic gas in the anesthesia machine piping and breathing circuit
needs to be flushed out in anticipation of reconnecting the patient
to the anesthesia machine. If the gas in the opened O2 cylinder is O2,
then FiO2 will increase as the gas from the O2 cylinder replaces the
hypoxic gas. A circuit disconnected at the Y-piece also provides for
faster flushing of the hypoxic gas from the circuit. An increase in
FiO2 when using gas from the O2 cylinder confirms that that gas supplied
by the O2 pipeline is hypoxic. Now that the gas in the O2 cylinder is
confirmed to be O2, if the patient requires O2-enriched air during ventilation
with an SIRB or Mapleson system and a stand-alone O2 cylinder is not
available, supplemental O2 from an auxiliary O2 flow meter or common
gas outlet installed on the anesthesia machine may now be used if the
O2 pipeline remains disconnected from the anesthesia machine. The plumbing
leading up to auxiliary O2 outlets may be flushed of residual hypoxic
gas by opening them momentarily prior to connection to the SIRB or Mapleson
system. If time allows, a further confirmatory test may be performed
by verifying that the measured FiO2 corresponds to the flow meter settings
or set FiO2. Fuel (galvanic) cell O2 sensors are common in anesthesia
machines and have a slow response time compared to the faster paramagnetic
O2 analyzers. The user should be patient when waiting for the FiO2 to
increase with these slow O2 sensors.
(g) Alert all other locations supplied by the same
central oxygen supply
Simultaneous patient deaths in multiple anesthetizing locations supplied
by the same oxygen central supply have been reported. Early detection
in one room and prompt communication of the problem to other sites using
oxygen from the same O2 central supply may save lives and prevent multiple
deaths. This proposed algorithm points out the need to develop procedures
for alerting all locations (not limited to anesthetizing locations)
in a hospital or institution supplied by a given O2 central supply system
as well as immediately informing the O2 manufacturer so that other institutions
supplied by the same manufacturer or plant may be warned in a timely
manner.
(h) Reconnect breathing circuit to patient, resume
ventilation (mechanical or spontaneous) using O2 from the O2 cylinder
and order more O2 tanks
Volatile anesthetics cannot be safely delivered while ventilating with
an SIRB or Mapleson system. If the surgical case is still in progress,
reconnect the patient to the machine supplied with O2 from the O2 cylinder
and resume inhalational anesthesia.
(i) If extra O2 cylinders are not readily available
and patient was previously being mechanically ventilated, ventilate
manually using the reservoir bag to save the oxygen used as ventilator
drive gas
As discussed above, O2 makes up all or part of the drive gas for gas-driven
anesthesia ventilators. To eliminate the oxygen consumed by gas-driven
anesthesia ventilators and make the finite amount of gas in a back-up
O2 E-cylinder (660 L when full at 2,000 psig) last longer, the clinician
should resort to manual ventilation if the expected case duration is
close to or exceeds the calculated time to exhaustion of the O2 E-cylinder.
The time to exhaustion is calculated by dividing the residual O2 volume
in the cylinder by the rate of consumption of O2. Residual volume in
liters (L) in an E-cylinder is calculated by dividing the cylinder pressure
in psig by 2000 psig and multiplying by 660 L. For example, 1,000 psig
represents (1000/2000) * 660 = 330 L. The rate of consumption of O2
during mechanical ventilation is the sum of the O2 flow meter setting
and the minute ventilation (tidal volume in L x respiratory rate in
breaths/min). For example, if FGF is 0.5 L/min O2 and 1.0 L/min N2O
and VT is 0.7 L and RR is 10 bpm, then the minute volume is 7 L/min
and the total O2 consumption is 7.5 L/min. The expected time to exhaustion
is thus approximately 330 L/7.5 L/min = 44 minutes, ignoring the gas
sampled by the gas analyzer and leaks.